The REMOTE 5104PN meets the stringent requirements of today’s aseptic processing applications. Incorporating the latest laser optical particle counting technology, Lighthouse designed the REMOTE 5104PN for continuous trouble-free operation.

 

     The REMOTE 5104PN uses a built-in vacuum pump and flow control system to insure accurate and reliable sensor performance.

 

     With a sensitivity of 0.5 micron at a controlled flow rate of 1.0 CFM, the REMOTE 5104PN provides continuous real-time data at a cost-effective price per point.

 

     The REMOTE 5104PN integrates seamlessly into large facility monitoring and management systems and transfers up to 4 channels of simultaneous particle count data using RS-485 MODBUS or Ethernet MODBUS TCP.

 

Designed and built by Lighthouse, a name you can trust. 

 

Features

  • Channel sizes: 0.5 - 5.0μm for GMP Compliance
  • 1.0 CFM (28.3 LPM) Flow Rate
  • Internal Vacuum Pump with Flow Control
  • Up to 2 Channels of Simultaneous Data
  • Meets ISO 21501-4 Standards
  • NEMA Style 316L Stainless Steel Enclosure or
  • NEMA Style Powder Coated Steel Enclosure
  • LED Status Indicators
  • Interfaces to Facility Monitoring Systems
  • RS-485 MODBUS ASCII or Ethernet MODBUS TCP Output
  • 120VAC or 24VDC, 100W Models Available

Benefits

  • 2 Year Warranty
  • Cost Effective Price per Point
  • Easy System Integration
  • International Sales and Service Support
  • Low Cost of Ownership

Applications

  • Aseptic Process Monitoring
  • Isolator Monitoring
  • Aerospace
  • Chemical Manufacturing
  • Equipment Monitoring
  • Pharmaceutical Packaging and Manufacturing
  • Trend Analysis
  • Cleanroom Laundries
  • Cleanroom Packaging
  • Semiconductor
  • Hard Disk Drive Manufacturing

Specifications

Size Range: 0.5 - 25μm
Channel Sizes:

Standard: 0.5, 5.0μm
Standard: 0.5, 10.0μm

Counting Efficiency: 50% @ 0.5μm, 100% for particles >0.75μm (per ISO 21501-4)
Flow Rate: 1.0 CFM (28.3 LPM)
Laser Source: Laser Diode
Zero Count Level: < 1 count / 5 minutes (per ISO 21501-4)
Concentration Limits: 500,000 particles / ft3 @ 5% coincidence loss
Calibration: NIST Traceable
Data Storage Rotating Buffer, 2000 Records
Sample Output: Internally filtered to HEPA standards (>99.97% @ 0.3μm)
Vacuum Source: Internal Pump
Air Flow: Internally Monitored and Controlled
Communication Modes: MODBUS: RS-485; Optional Ethernet: Ethernet MODBUS TCP
LED Indicators: Power, Flow, Service, Sampling
I/O Connection: RJ-45
Supporting Software: Lighthouse Monitoring System; LMS XChange; LMS Express, Express RT
Enclosure: NEMA Style 316L Stainless Steel or Powder Coated Steel
Power: 110 / 220 VAC, 50-60Hz or 24VDC 100W
Dimensions: 10" (w) x 12" (h) x 6.9" (d) [25.4 x 30.4 x 14.9 cm]
Weight: AC Model, 20 lbs (9 kg); DC Model, 19.5 lbs (8.85 kg)

Conditions

Operating: 50°F to 104°F (10°C to 40°C), Humidity 20% to 95% non-condensing
Storage: 14°F to 122°F (-10°C to 50°C), Humidity up to 98% non-condensing

Accessories

Included:

Operating Manual on CD; AC or DC Power Input Connector Kit; Industrial RJ45 Kit; Fuse(s)

Optional:

Printed Operating Manual; Cabling; Data Cable; Sample Tubing; 24VDC

100W Power Adapter (DC Model)